Contact an Account ManagerSeparation of 20 PAHs on NUCLEODUR® C18 PAH
Summary
Type: HPLC
Application #: 125710
Catalog#: 760784.40 - MN HPLC Column, Nucleodur C18 PAH, 110A, 3um, 4x125mm
Detection:
Conditions: Eluent A: acetonitrile
Eluent B: methanol / water (70:30, v/v)
Gradient: 15 % A for 2 min, to 20 % A in 3 min, to 50 % A in 3.5 min, to 65 % A
in 1.5 min, hold for 3 min, to 100 % A in 2 min, hold for 7 min
Flow rate: 0.55 mL/min
Temperature: 25 °C
Injection volume: 5.0 µL
Authors: HPLC department
Source: Macherey-Nagel, Germany 2012